Property
Description
Setting resolution
1
Available in modes
•
PCV+A
•
PSV
•
S
•
S/T
4.2.10
Expiratory Trigger (Exp.Trigger)
The Expiratory Trigger setting defines the moment when the ventilator will cycle from
the inspiratory to the expiratory phase.
Property
Description
Value range
Auto, 1 to 9, corresponding to 10% to 90% of peak flow
Default value
Auto
Setting resolution
1
Available in modes
•
PSV
•
S
•
S/T
4.2.11
Rise Time
The Rise Time setting controls the speed of the pressure increase from start of inspira-
tion to set inspiratory pressure.
A low setting will give a faster pressure increase and therefore a longer plateau at the set
inspiratory pressure. A high setting will give a slow increase and therefore a shorter
plateau.
Property
Description
Value range
Auto, 1 to 9
Default value
Auto
Setting resolution
1
Available in modes
•
PCV+A
•
PSV
•
S
•
S/T
•
T