7.4.10
Internal Function Failure
Property
Description
Alarm text
Int. Function Failure(01–18)
Priority
High
Alarm condition
Failure of internal function that prevents treatment or normal
operation of the ventilator.
The error code that follows the alarm text indicates the kind
of function failure.
Reset criteria
Correct function is restored.
Ventilator action
The ventilator will stop the treatment.
Action to take
Restart the Vivo 3.
If the alarm persists or reoccurs: Take a note of the error
code and contact your supplier of the Vivo 3 .
108 Alarms
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2