Activate the heated circuit
1
Press the
Setup
button twice.
⇒
The second setup page is now displayed, with heated circuit settings.
2
Press the
Up arrow
button until the setting
Heated Circuit
is selected.
3
Press the
Right arrow
button to turn the heated circuit on.
⇒
The value is shifted to On and the heated circuit symbol is lit on the display.
⇒
The setting for Heated Circuit Level is displayed.
4
Press the
Down arrow
to select the
Heated Circuit Level
setting.
5
Press the
Left arrow
button to decrease the heating or the
Right arrow
button to
increase the heating.
6.7.3
Using the SpO
2
Sensor
The SpO
2
module (consisting of a SpO
2
sensor, an electronic unit) is intended to meas-
ure functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate.
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2