Mve (Minute Volume, expiratory)
is calculated as the Tidal Volume multiplied with the
Total Breath Rate.
•
Vte
Vte (Tidal volume, expiratory)
is the expired tidal volume for each breath.
•
% in TgV
Percentage of breaths within the target volume window.
•
Rise Time
Rise Time
displays the duration of the pressure increase during the inspiration phase.
•
SpO2
SpO
2
(Saturation of Peripheral Oxygen)
displays the patient
’
s oxygen saturation. as meas-
ured by the SpO
2
module.
This value is only displayed if the SpO
2
module is connected.
Displayed ranges: According to the manufacturer
’
s specifications.
•
Pulse Rate
Pulse Rate
displays the patients pulse rate as measured by the SpO2 module.
This value is only displayed if the SpO
2
module is connected.
Displayed ranges: According to the manufacturer
’
s specifications.
6.6.2
Treatment Values Monitored by External Equipment
Monitoring of Expiratory CO
2
For monitoring expiratory CO
2
, an external monitoring device shall be connected to the
patient circuit.
The device shall comply with ISO 80601-2-55.
Monitoring of Oxygen Saturation
For monitoring the patient air oxygen saturation, an external monitoring device shall be
connected to the patient circuit.
The device shall comply with ISO 80601-2-55 and have a high oxygen level alarm.
6.7
Using Accessories
This section describes how to use accessories provided by Breas Medical.
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual
Содержание Vivo 3
Страница 1: ...DOC 007232 EN EU Y 2 2019 09 03 Vivo 3 Clinician s Manual...
Страница 2: ......
Страница 8: ......
Страница 136: ......
Страница 137: ...Appendix Appendices 137 Doc 007232 Y 2 Vivo 3 Clinician s Manual Appendices...
Страница 140: ......