8.4.2.2 Low Pressure Alarm and the Disconnection Alarm
1. Start the treatment
2. Disconnect the patient circuit.
3. Wait 15 seconds.
⇒
The Low Pressure Alarm and/or the Disconnection Alarm will be given.
8.4.2.3 Low Breath Rate Alarm
1
Connect the Vivo 3 to a patient circuit.
2
Set the Low Breath alarm to 10 bpm.
3
Adjust the Vivo 3 settings as follows:
•
Ventilation Mode: T
•
Target Volume: Off
•
IPAP: 10 cmH
2
O
•
EPAP: 5 cmH
2
O.
•
Breath Rate: 12 bpm
4
Start the treatment on the Vivo 3 device.
5
Wait until the Low Breath Rate alarm is activated.
6
Stop the treatment. The test is completed.
8.4.2.4 High EPAP Alarm
1
Connect the Vivo 3 patient circuit to a test lung and a CPAP device.
2
Set the CPAP device treatment pressure to 10 cmH
2
O.
3
Adjust the Vivo 3 settings as follows:
Setting
Value
Ventilation Mode
PCV+A
Target Volume
Off
IPAP
15 cmH
2
O
EPAP
5 cmH
2
O
Breath Rate
12 bpm
Insp. Time
1.5 s
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2