7.3.10
High EPAP Alarm
Property
Description
Alarm text
High EPAP
Priority
Medium
Alarm condition
A High EPAP alarm will be given when EPAP has gone 30%
above the set value for 3 breaths.
Possible cause
•
Blocked leakage port.
•
Too short expiratory time.
•
Changes in airway resistance and or compliance.
Reset criteria
EPAP has gone below the alarm limit (lower than 30% above
the set value).
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
On
•
Off
7.3.11
Low EPAP Alarm
Property
Description
Alarm text
Low EPAP
Priority
Medium
Alarm condition
A Low EPAP alarm will be given when EPAP has gone 30%
below the set value for 3 breaths.
Possible cause
•
Excessive leakage.
Reset criteria
EPAP has gone above the alarm limit (higher than 30%
below the set value).
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
On
•
Off
Alarms 101
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual