4.2.14
Min Pressure
The Min pressure setting is only used when Target Volume is activated.
Min pressure defines the lower pressure limit down to where the Vivo 3 can decrease
the pressure to maintain the set Target Volume. If the actual volume is above Target
Volume at Min pressure, the Vivo 3 will continue to ventilate at this Min pressure
setting.
Property
Description
Value range
4 cmH
2
O to 20 cmH
2
O.
Default value
Same as the value for
IPAP
Setting resolution
Below 10 cmH
2
O: 0.5 cmH
2
O
Above 10 cmH
2
O: 1.0 cmH
2
O
Available in modes
•
PCV+A
•
PSV
•
S
•
S/T
•
T
4.2.15
Ramp Up
The ramp up setting defines a ramp time for increasing the airway pressures.
At the beginning of Ramp Up time, IPAP will start at 2 cmH
2
O above the initially set
EPAP.
Property
Description
Value range
•
Off
•
10 min to 60 min
Default value
Off
Setting resolution
10 min
Available in modes
•
PCV+A
•
PSV
•
S
•
S/T
•
T
•
CPAP
Treatment Functions and Settings 59
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual