Maximum Flow Rate at Continuous Pressure
Pressure (hPa)
Flow Rate (l/min)
4
76
8
105
12
158
16
157
20
149
Maximum Pressure
Property
Value
Maximum pressure limit in the event
of fault condition
60 cmH
2
O
CPAP Maximum Dynamic Pressure Variations
Device with standard tube / Device with worst case VBS (external humidifier with 22
mm tube and bacterial filter)
Pressure [cmH2O]
10 bpm
15 bpm
20 bpm
4
0.3/1.0
0.5/1.2
0.7/1.6
8
0.3/1.0
0.5/1.2
0.7/1.6
12
0.3/1.0
0.5/1.2
0.7/1.6
16
0.3/1.0
0.5/1.2
0.7/1.6
20
0.3/1.0
0.5/1.2
0.7/1.6
Inlet Air Filter Specification
Specification for Air inlet filter, white (part no 007202):
Performance
Characteristic
Value
Tested in Accordance with
NaCl Penetration
<7.35 %
BS EN 13274
–
7 NaCl at 16 cm/sec
Air flow resist-
ance (pressure
drop) flow
12.5 Pa typical
20 Pa max
at 82.5 l/min (100 cm
2
)
0.3 Pa max
at 0.2 cm/sec
Doc. 007232 Y-2