With Humidifier
1
2
3
4
5
6
Number
Description
1
Air inlet (Ambient room air intake)
2
Blower
3
Sensors
4
Humidifier
5
Air outlet (Breathing air to the patient)
6
Cartridge heater
The humidification output is controlled by the ventilator by regulating the power to the
cartridge heater in the humidifier.
9.5
Technical Data
Noise Levels
Property
Value
Static sound pressure level
< 30 dB(A) @ 10 cmH2O, 4 mm leak.
Measured at a distance of 1 meter, according
to standard.
Static sound pressure level, stand by
mode
< 22 dB(A) SPL.
Measured at a distance of 1 meter, according
to standard.
Maximum sound pressure level
40 dB(A) SPL according to applicable
standards.
Measured at a distance of 1 meter, according
to standard.
Maximum sound power level
48 dB (A) according to applicable standards.
Measured at a distance of 1 meter, according
to standard.
Vivo 3 Clinician
’
s Manual
Doc. 007232 Y-2