Rev. 5.01
35
Note
When external power is not supplied, it takes about 15 seconds for the battery
status display to reflect the actual remaining battery capacity.
Battery information
⚫
Battery specification: Li-ion, 10.8V
3BL335-BIO-S (3250mAh) or 031PpTC3(3ICR19/65) (2150mAh) (BM3Vet Elite, BM5Vet Elite
standard)
6BL335-BIO-S (6500mAh) (BM7Vet Elite standard, BM3Vet Elite/BM5Vet Elite optional)
⚫
Battery charging time:
3BL335-BIO-S: More than 3 hours
6BL335-BIO-S: More than 5 hours
⚫
Continuous battery usage time when fully charged:
BM3Vet Elite
- 3BL335-BIO-S: about 2 hours
- 6BL335-BIO-S: about 4 hours
BM5Vet Elite
- 3BL335-BIO-S: about 2 hours
- 6BL335-BIO-S: about 4 hours
BM7Vet Elite
- 6BL335-BIO-S: about 2 hours
* (measured every 15 minutes NiBP with SpO2 and ECG)
Note
The Lithium-Ion battery is a rechargeable battery containing Lithium-Ion cells.
Each battery contains an integrated electronic fuel gauge and a safety
protection circuit.
Warning
Older or defective batteries will have significantly reduced capacity or
operating time.
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...