Electrostatic Discharge
Immunity (ESD) IEC 61000-
4-2
±
8 kV/Contact
±2, ±4, ±8, ±15 kV/Air
±
8 kV/Contact
±2, ±4, ±8,
±15 kV/Air
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least 30%.
Radiated RF
Electromagnetic Field
Immunity IEC 61000-4-3
3 V/m 80 MHz - 2.7
GHz 80% AM at 1 kHz
3 V/m 80 MHz - 2.7
GHz 80% AM at 1 kHz
BM3Vet Elite, BM5Vet
Elite, BM7Vet Elite are
suitable to use in
professional healthcare
environment.
RF communication
equipment is used no
closer than 30 cm to
any part of the BM3Vet
Elite, BM5Vet Elite,
BM7Vet Elite, including
cables specified by
Bionet
Immunity to Proximity
Fields from RF wireless
Communication s
Equipment IEC 61000-4-3
28 V/m Max. 3855785
MHz in according to
table 9 in IEC 606011-2
28 V/m Max. 3855785
MHz in according to
table 9 in IEC 606011-2
Mains power quality
should be that of a
typical commercial or
hospital environment.
Electrical Fast
Transient/Burst Immunity
IEC 61000-4-4
±
2 kV, 100 kHz
repetition frequency
±
2 kV, 100 kHz
repetition frequency
The quality of supplied
power should be
suitable for general
business site or
hospital environment.
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...