Rev. 5.01
83
⚫
ECG signal degradation
⚫
Excessive sweating of the animal
⚫
Animal's skin irritation
There are a variety of reusable and disposable electrodes available. Choose the electrode that best
fits your monitoring situation. Bionet recommends Ag / AgCl disposable electrodes.
If you are using an electrode with a gel beforehand, make sure that the electrode is sufficiently
gelled. Never use this product if the disposable electrode has expired or the gel is dry.
Determine the electrode location that will provide the best ECG in the configuration (P-wave and
T-wave amplitudes should not exceed 1/3 of the QRS amplitude). Choose a flat, non-muscular
location to maximize contact with the electrodes and minimize muscle fatigue.
Avoid joints or bony protrusions. When choosing a location for electrode placement, consider the
following special conditions:
⚫
Surgery - Place electrodes as far away from the surgical site as possible.
⚫
Burn animal - use sterile electrodes. Thoroughly clean the equipment. Follow hospital
infection control procedures.
Use a waterproof tape (about 2 inches wide) or Steri-Drape to secure the electrode. Protect from
liquids. Make a small loop from the lead wire just below the connection and secure with tape.
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...