Rev. 5.01
133
hardware error may exist.
CO2 OUT OF
RANGE
The value being calculated is greater
than the upper CO2
limit( 150mmHg )
If error persists, perform a zero.
CHECK AIRWAY
ADAPTER
Usually caused when the airway
adapter is removed from the
Capnostat or when there is an
optical blockage on the windows of
the airway adapter. May also be
caused by failure to perform
Capnostat zero to when adapter
type is changed.
To clear, clean airway adapter if
mucus or moisture is seen. If the
adapter is clean, perform a
Capnostat zero.
MODULE OFF
It occurs when the equipment and
module are separated. Message
output
Verify module connections
Service request
CO2 measurement failure
CO2 value is not output, or numerical error.
Troubleshoot procedure
1.
Check the connection between the main unit and the module.
2.
Check the module line connection with the filter line or airway.
3.
Replace filter line or airway.
4.
Service Request
Note
In the following monitoring conditions, the measured values may be
inaccurate. Read the measured values carefully.
1.
When using this in an environment of using nitrous oxide gas of high
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...