Rev. 5.01
181
IBP Performance (BM5Vet Elite/BM7Vet Elite Only)
Channels
2
Measurement Range
-50 to 300mmHg
Accuracy
4 % of reading or ± 4 mmHg
, whichever is greater
Pulse Rate
Measurement Range
0 to 300bpm
Zero Balancing
Range:
±
200mmHg
Accuracy:
±
1mmHg
Drift:
±
1mmHg over 24hours
Transducer Sensitivity
5µV/
V/mmHg
Arrhythmia Alarm Level and Alarm Detection On/Off
Type
Arrythmia
Alarm Level
Alarm On/Off
High
Medium
Low
Messge
On
Off
OFF
ASYSTOL
●
●
VTAC
●
●
VTACVFIB
●
●
BIGEMINY
●
●
TRIGEMINY
●
●
ACCVENT
●
●
COUPLET
●
●
IRREGULAR
●
●
PAUSE
●
●
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...