Trouble Shooting
Problem
Solution
Inaccurate heart rate
and/or false asystole.
Check ECG signal from patient:
1.
Check/adjust lead placement.
2.
Check/perform skin preparation.
3.
Check/replace electrodes.
Check amplitude of ECG waveform:
1.
Select ECG parameter label.
2.
Select DISPLAY LEAD,
3.
Scroll through all ECG leads and check for 0.5mV
amplitude at normal (1X) size. (at least 0.5mV amplitude is
required for QRS detection.) for borderline signals, validate
on a graph.
4.
If amplitudes are low, electrodes may need to be
repositioned or replaced.
False ventricular fibrillation
occurrence
Check ECG signal from patient: (the chest lead may exhibit polarity
changes which may occasionally cause an inaccurate call.)
1.
Check/adjust electrode placement.
2.
Check the electrode attachment to the skin and attach it
correctly.
3.
Check the condition of the electrode and replace it if
necessary. (if chest lead is a problem, move the chest
electrode to another chest position or leg position.)
Inaccurate pacemaker
detection
Use pacemaker processing:
1.
Select ECG parameter label.
2.
Display the lead of ECG with the greatest amplitude in the
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...