Rev. 5.01
129
A-2.
Display option
EtCO2 parameter wave display setup
A-2-1.
Sweep speed
Waveform sweep speed setup
6.25mm/s,
12.5mm/s,
25mm/s
A-2-2.
Scale
Display waveform scale setup.
The selectable value is the maximum
pressure range shown in the waveform.
When you select a range value, the
selected pressure range value is
displayed below the dotted line above
the two dotted lines in the left middle
of the wave window.
40mmHg (5.3 vol%)
50mmHg (6.6 vol%)
60mmHg (7.9 vol%)
80mmHg (10.5 vol%)
100mmHg(13.2vol%)
150mmHg(19.7vol%)
A-2-3.
Fill
Choose whether to fill the waveform
inside
On / Off
A-2-4.
Gas pressure unit
Choose gas pressure unit
mmHg
kPa
vol%
A-2-5.
Use one gas unit
Choose to set the pressure unit for
each gas type
Unit setting menu by gas type appears
when off.
On / Off
A-3.
Apnea detect
APNEA detection menu
On / Off
A-4.
Module information
Module information menu
A-4-1.
Sensor PN
The sensor part number
PNXXXXXX
A-4-2.
OEM ID
The id is a 7-bit identifier which is set
at the factory to a unique value for
each OEM.
0X01
A-4-3.
sensor SN
The serial number of the module
A-4-4.
H/W version
The hardware version number of the
module
A-4-5.
Total usage time
Total use time of the module.
A-4-6.
Last zero time
This is the total time that has elapsed
with the sensor in service the last zero.
Min. display
A-4-7.
Pump total time
This is the total time the pump has
Min. display
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...