Rev. 5.01
95
COUPLET
1)
Occurs when two ventricular beats are detected and have non-ventricular
beats before and after the couplet. The coupling interval must be less than
600 milliseconds.
IRREGULAR
1)
Occurs when six consecutive normal R-to-R intervals vary by 100
milliseconds or more.
PAUSE
1)
Occurs when the interval between two consecutive beats exceeds three
seconds.
PVC
1)
Isolated premature ventricular complexes occur when a premature ventricular
beat is detected and has non-ventricular beats before and after.
R ON T
1)
Occurs when a ventricular complex is detected within the repolarization
period of a non-ventricular beat.
TRIGEMINY
1)
Occurs when two or more trigeminal cycles (a ventricular beat followed by
two non-ventricular beats) are detected.
V BRADY
1)
Ventricular bradycardia occurs when a run of three or more ventricular
beats is detected with an average heart rate that is less than or equal to 50
beats per minute.
1)
: BM5Vet Elite, BM7Vet Elite only
Note
The Brady limit matches the low heart rate limit. If the low heart rate limit is
changed, the Brady limit changes.
Arrhythmia Settings
Menu
Description
Available settings
A. ECG menu
A-1.
Arrhythmia
Arrhythmia parameter alarm setup
menu
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...