Rev. 5.01
139
base line. And accord the wave line with the data.
5.
Set the data as ‘
0
’ on the parameter screen.
6.
Check if
Zero reference
is carried out. (Check the pressure parameter on the message
window.)
7.
Close the venting stopcock on the air side.
8.
Open the transducer stopcock on the animal side. The pressure value should be displayed
on the pressure parameter screen in a few seconds.
Disconnected alarm
Disconn alarm menu
will be displayed when measurement label is set for ART, FEM and UAP.
This function will be activated upon the following two conditions.
1.
In case
Mean pressure
is not higher than 25mmHg.
2.
In case the
Disconnect alarm
is set ‘
ON’
.
Medium alarming sound will be generated when the
Dissconnected alarm
is activated, and the
alarming message “
DISCONNECTED
” will be displayed on the parameter screen.
CAL. TRANSDUC:
A function to adjust a Transducer error on the monitor
A function to adjust an error value based on the other index manometer.
How to Adjust
1.
Select a menu by pressing the knob switch key.
2.
Measure blood pressure along with another index manometer.
3.
Compare the measured values of ‘mmHg’ for both manometers.
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...