Rev. 5.01
131
gasses where the temperature is often
room temperature or below.
A-5-6.
O2 compensation
Use this setting to correct for the
compensation of the gas mixture
administered to the animal.
A-5-7.
Anesthetic agent
Anesthetic agent is ignored when the
balance gas is set to helium.
A-5-8.
Zero type
This function is used to initiate a
When performing a zero on room air,
this setting should be set to room air
(the default). Only change to nitrogen
(N2) when performing a zero on 100%
N2 gas; this is provided for use in a
laboratory environment.
Room air
N2
B-1.
Zeroing
This function is used to initiate a
Capnostat zero.
A zero is used to correct for
differences in airway adapter types.
The Capnostat zero must be
performed free of any CO2
1.
Set the Host to the zeroing
function.
2.
Connect the CAPNOSTAT 5
CO2 Sensor.
3.
Place the CAPNOSTAT 5 CO2
Sensor onto a clean and dry
CO2 adapter that is exposed
to room air and away from all
sources of CO2, including the
ventilator, the
animal’s
breath
and your own.
Start the adapter zero. The maximum
time for a CAPNOSTAT zero is
40seconds. The typical time for a zero
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...