Portable and mobile RF communication equipment can affect medical electrical
equipment.
Cables and accessories not specified in the user guide are not certified. Using
other cables and / or accessories may adversely affect safety, performance, and
electromagnetic compatibility (increased electromagnetic emissions and
reduced immunity).
⚫
This equipment should not be used near or on top of other equipment.
If you need to use it on its side or stacked, you should observe the
equipment to make sure it works properly within your configuration.
⚫
This veterinary multiparameter monitoring device communicates over
a 2.4 GHz 802.11b / g wireless network. Other equipment may
interfere with data reception on this wireless network. This is also true
if the equipment complies with the CISPR emission requirements.
When using veterinary multiparameter monitoring equipment to
communicate over a wireless network, be sure to check that it is
compatible with existing or new wireless systems (eg, cell phones,
pager systems, cordless phones, etc.). For example, a Bluetooth-
compliant device using the 2.4 GHz frequency band may interfere with
the wireless communication of the Veterinary Multiparameter Monitor.
For more information on wireless deployment, please contact your
Bionet representative.
⚫
Low amplitude signals such as EEG and ECG are particularly sensitive
to interference from electromagnetic energy. This equipment complies
with the tests listed at the bottom but does not guarantee complete
operation. The "quiet" electrical environment is better. In general, the
greater the distance between electrical equipment, the lower the
likelihood of interference.
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...