Rev. 5.01
115
1.
Check for proper cuff size.
2.
Check for residual air left in the cuff from a previous measurement.
3.
Make sure cuff is not too tight or too loose.
4.
Make sure cuff and heart are at same level, otherwise hydrostatic
pressure will offset the NIBP value.
5.
minimize patient movement during measurement.
6.
Check for leak in cuff or tubing.
7.
Patient may have a weak pulse.
Measurement Limitations
The measurement may be inaccurate or impossible:
⚫
Measurements are impossible with heart rate extremes of less than 40 bpm or greater than
300 bpm, or if the patient is on a heart-lung machine.
⚫
With excessive and continuous patient movement such as shivering or convulsions
⚫
if a regular arterial pressure pulse is hard to detect
⚫
With cardiac arrhythmias
⚫
With rapid blood pressure changes
⚫
With severe shock or hypothermia that reduces blood flow to the peripheries
⚫
With obesity, where a thick layer of fat surrounding a limb dampens the oscillations coming
from the artery
⚫
On an edematous extremity.
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...