Rev. 5.01
105
Note
SpO2 wave size changes automatically
Signal and Data Validity
It is extremely important to determine that the probe is attached to the animal correctly and the
data is verifiable. To make this determination, three indications from the monitor are of assistance
—
signal strength bar, quality of the SPO2 waveform, and the stability of the SPO2 values. It is critical
to observe all three indications simultaneously when ascertaining signal and data validity.
Signal Strength Bar
The signal strength bar is displayed within the SPO2 values window. This bar consists of 10 blocks
set depending on the strength of the signal. Proper environmental conditions and probe attachment
will help to ensure a strong signal.
Quality of SPO2 Waveform
Under normal conditions, the SPO2 waveform corresponds to (but is not proportional to) the arterial
pressure waveform. The typical SPO2 waveform indicates not only a good waveform, but helps the
user find a probe placement with the least noise spikes present. The figure below represents an
SPO2 waveform of good quality.
Good Quality SPO2 Waveform
If noise (artifact) is seen on the waveform because of poor probe placement, the photo detector
may not be flush with the tissue. Check that the probe is secured and the tissue sample is not too
thick. Pulse rate is determined from the SPO
2
waveform which can be disrupted by a cough or other
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...