Rev. 5.01
49
B-3.
IBP
1)
B-3-1.
IBP1 calibration
1)
Perform 100mmHg
pressure calibration and
select menu
B-3-2.
IBP2 calibration
1)
Perform 100mmHg
pressure calibration and
select menu
C.
Screen calibration
Perform touch screen calibration point input
D.
Maker services
D-1.
MAC address editing
Enter a unique address
for the device
D-2.
Factory reset
Reset menu for setting the device
to factory default state
Perform factory reset
E. SW upgrade
Software Upgrade menu
F. SW license
Software License menu
1)
: BM5Vet Elite, BM7Vet Elite only
Parameter color
Selectable colors
green, light blue, magenta, yellow, blue, sky blue, white, coral, scarlet, purple, orange, pale green,
pink, pale yellow
Parameter
Default color
Parameter
Default color
ECG (ST)
Green
SpO2
Light Blue
RESP
Yellow
NIBP
Magenta
TEMP
Green
ETCO2
Yellow
IBP1
1)
Scarlet
IBP2
1)
Light Blue
1)
: BM5Vet Elite, BM7Vet Elite only
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...