Electromagnetic Compatibility
The monitor has been designed and tested for compliance with current regulatory standards as to
its capacity to limit electromagnetic emission (EMI), and also as to its ability to block the effects
of EMI from external sources.
The monitor complies with the following standards pertaining to EMI emissions and susceptibility:
EN60601-1-2.
To reduce possible problems caused by electromagnetic interference, we recommend the
following:
⚫
Use only Bionet approved accessories.
⚫
Ensure that other products used in areas where veterinary multiparamter monitoring and
life support is used comply to accepted emissions standards (CISPR 11, Class A).
⚫
Try to maximize the distance between electro-medical devices. High-power equipment
related to electrical simulators, electrosurgical instruments, and radiators (X-ray machines)
as well as evoked potential devices may cause monitor interference.
⚫
Strictly limit exposure and access to portable radio frequency sources (e.g., cellular
phones and radio transmitters). Be aware that portable phones may periodically transmit
even when in standby mode.
⚫
Maintain good cable management. Do not route cables over electrical equipment. Do not
intertwine cables.
⚫
Ensure all electrical maintenance is performed by qualified personnel.
Caution
Infectious devices and parts must be sanitized and cleaned before disposal.
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...