Your system administrator can change the alarm priority level for the nurse call signal.
If the priority level is set to
High,
only high-priority alarms will sound on the nurse call system.
Note
Audio Paused and Audio Off modes only stop the audible alarm sound and
touch or key sound is activated always.
To adjust the Touch or Key Sound, use the Key Sound menu in Setup.
Alarm Settings
Main menu
Sub menu
A.
Alarm setup
A-1.
Parameter alarm limit
A-2
. Arrhythmia alarm condition
A-3.
System alarm condition
A-4.
Alarm parameter
A-5.
Nurse call
B.
Alarm review
Sub menu
Description
Available settings
A. Alarm setup menu
A-1.
Parameter alarm limit
All parameter alarm, level, activate
Setup menu
A-2.
Arrhythmia alarm
condition
ASYSTOLE, VTAC, VTAC /VFIB
Only for BM5Vet Elite/BM7Vet Elite:
BIGEMINY, TRIGEMINY, ACCVENT,
COUPLET, IRREGULAR, PAUSE,
RONT, VBRADY, SHORTRUN, PVC
A-3.
System alarm condition
Low battery
A-4.
Alarm parameter
Alarm settings menu
A-4-1.
Alarm volume
The volume can be changed from OFF
to 10% to 100%.
Off, 10~ 100%
A-4-2.
Alarm pause time
No sound for 5minutes, Release on
alarm again
1,2,3,5,10,15min
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...