User and Programming Manual
36 XDS
Series
as a percentage of voltage range. Allowable range is 0 to 110%. Bus operation also allows the
OVP mode to be turned on (enabled) or off (disabled).
If the output voltage exceeds the set trip level and the OVP mode is enabled, the power supply
will generate a -300, Device specific error and turn off the output. The mode is enabled with the
VOLT:PROT:STAT command. The OVP mode provides another level of protection in addition to
the volt:high user setting which prevents programming commands or front panel inputs above this
level to be accepted.