User and Programming Manual
XDS Series
103
11.7.4 *IDN?
This query requests the power supply to identify itself. It returns the data in four fields separated by
commas. Note that any installed options such as –IF are not reflected in the *IDN? query response.
The presence of these options that may affect the power supply’s capabilities must be determined by
using other queries.
Query Syntax
*IDN?
Returned Parameters
<AARD>
Field
Information
AX
Manufacturer
xx-xx
Model number, Volt - Current
nnnnn
Serial number or 0
Rev. x.x
Revision level of firmware
Example "AX,XDS80-63,12345,Rev
1.0"
11.7.5 *RCL
This command restores the power supply to a state that was previously stored in memory with a
*SAV command to the specified location. The XDS Series offers eight non-volatile memory locations
for storing instrument setups.
Command Syntax
*RCL <NRf>
Parameters
0 through 7
Example *RCL
3
Related Commands
*RST
*SAV
WARNING:
Recalling a previously stored state may place hazardous voltages at the
power supply output if the output state saved is ON.