User and Programming Manual
94
XDS
Series
[SOURce:]VOLTage:PROTection[:LEVel]
This command programs the OVP trip level of the power supply.
The OVP trip level is set as a percentage of voltage range. Allowable range is 0 to 110%. If the
output voltage exceeds the set trip level and the OVP mode is enabled, the power supply will
generate a -300, Device specific error and turn off the output. The mode is enabled with the
VOLT:PROT:STAT command. The OVP mode provides another level of protection in addition to the
volt:high user setting which prevents programming commands or front panel inputs above this level
to be accepted. The OVP function is based on the measured output voltage rather than the
programmed output value.
Command Syntax
[SOURce:]VOLTage:PROTection[:LEVel] <NR1>
Parameters
percentage of V range
Unit %
*RST Value
110
Examples
VOLT:PROT 80
VOLT:PROT:LEV 25
Query Syntax
[SOURce:]VOLTage:PROTection[:LEVel]?
Returned Parameters
<NR1>
Related Commands
VOLT:PROT:STATE
[SOURce:]VOLTage:PROTection:STATe
This command enables or disables the OVP mode of the power supply.
The OVP mode state may be set using this command. If the output voltage exceeds the set trip level
and the OVP mode is enabled, the power supply will generate a -300, Device specific error and turn
off the output. If the OVP mode is disabled, nothing will happen. Valid parameters are 0 for OVP off
and 1 for OVP on.
Command Syntax
[SOURce:]VOLTage:PROTection:STATe <NR1>
Parameters
0 | 1
*RST Value
0
Examples
VOLT:PROT STAT 0
VOLT:PROT:STAT 1
Query Syntax
[SOURce:]VOLT:PROT:STAT?
Returned Parameters
<NR1>
Related Commands
VOLT:PROT[:LEV]
Содержание XDS 100-100
Страница 4: ...User and Programming Manual 4 XDS Series Electrical Safety Symbols Used in This Manual ...
Страница 27: ...User and Programming Manual XDS Series 27 Figure 3 5 Outline Drawing XDS DC Supply ...
Страница 41: ...User and Programming Manual XDS Series 41 Figure 5 4 Parallel Cable Interconnect Accessory ...
Страница 45: ...User and Programming Manual XDS Series 45 Figure 6 1 DC Power Supply Block Diagram IF option shown ...
Страница 69: ...User and Programming Manual XDS Series 69 Figure 8 3 Replaceable Fuse Locations ...