(3) Calibration method
[Procedure]
1.
To warm-up the reference oscillator for the MD8470A, turn on the
power of the MD8470A , leaving it for 24 hours.
2.
Input the standard frequency into the external reference input of the
frequency counter.
3.
Adjust the calibration trimmer for the crystal oscillator (
Freq Adj
on
the rear panel of the MD8470A) so that the frequency counter dis-
plays 10.000000 MHz
±
0.2 Hz.