3.4 When System Error Is Displayed
3.4.1
When temperature error is displayed
A system error display screen as shown in Fig.3.4.1-1 will be displayed if
the internal temperature of the MD8470A rises abnormally.
Fig. 3.4.1-1 System error (temperature) display screen
When the OK button on the screen is clicked or 60 seconds has elapsed af-
ter the error display screen is displayed, the MD8470A is forcibly set to
the standby state (Windows enters the turn off state).
When the MD8470A enters the standby state, check whether there is any
problem in the environment for the MD8470A.
•
Is the MD8470A being used in an environment out of the operating
temperature range (+5 to +40
°
C)?
•
Is the fan exhaust air opening located on the side of the MD8470A
plugged up?
The MD8470A may have failed if the error display screen is displayed
again, even when the operating environment has been improved. Con-
tact Anritsu Service and Sales office in this event.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание MD8470A
Страница 15: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 31: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 63: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 85: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 99: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 101: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 113: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...