Gen II
User Manual
18-3 – Monitor Magnitude and Phase
Chapter 18
Revision 2.01, Jan 2020
Use the [
] key to move to the “Input” field. Use the [
] key to toggle between the
selections in the “Input” field to select the output of “Input” field. The available
selections are: “A,” “B,” “C,” “D,” “A+B,” and “A-B.”
Use the [
] key to move down to the “Description” field. Enter a name from the keypad
in the description field. (Refer to Chapter 3, “Using the Gen II Analyzer”, if you are
unfamiliar with using the keypad.) The description field is optional.
Use the [
] key to move to the “Full Scale Vibration” field. The “Full Scale Vibration”
field is a toggle selection field. Use the [
] or [
] keys to toggle through the available
selections for the field. The available selections are: 0.01, 0.02, 0.05, 0.10, 0.20, 0.50,
1.00, 2.00, 5.00, 10.0, 20.0, 50.0, 60.0, 70.0, 80.0, 90.0, 100, 110, 120, 130, 140, 150,
200, 500, 1000, 2000, and 5000. This scale refers to the number of engineering units of
vibration amplitude specified in the “Vibration” field above. The full scale indicates the
maximum vibration amplitude you expect to acquire or the maximum amplitude of
interest. Choose the amplitude that will adequately display the full amplitude of any
specified limits as a minimum. If you do not expect amplitudes in excess of what would
normally be experienced for the equipment application, set this field as low as possible
while still allowing sufficient space to display the maximum limitations as stated above.
CAUTION
Sensors connected to Channel A and Channel B (or Channel C and Channel D in the Viper II) must
be of the same type. Using different sensors during this function will cause erroneous readings
Use the [
] key to move down to the “Sensor” field. Select a sensor type by using the
[
] key to toggle between selections.
Once all fields are filled press the [OK] key to begin acquiring vibration data. See
Chapter 24, “Reading Spectrum and Scales” for a detailed explanation of the information
contained on the acquisition screen.
When finished acquiring vibration data, press the [OK] key to stop.
Revision History:
1/21/2016
2.00: Major rewrite generalizing the manual to cover both the Viper II and Cobra II. This chapter was updated with
screenshots and corrections to minor typographical errors.
East
4/1/2020
2.01: Updated screenshots reflecting SP7 UI changes. Added information about Tach Filter fields.
Shively
Содержание Cobra II
Страница 91: ...Chapter 5 RESERVED Revision 1 00 Dec 2014 ...
Страница 92: ...Chapter 6 RESERVED Revision 1 00 Dec 2014 ...