Gen II
User Manual
13-21 – Transient Vibration Survey
Chapter 13
Revision 4.10, Apr 20
13.1.5.2.1.
Enter the “
Ref. Level:
”, Reference Level, typically in decibels (dB), and the “
Ref.
Frequency:
”, Reference Frequency, in Hertz (Hz), as found on the calibrator. Plug the
microphone into the correct size adapter in the top of the calibrator. Turn the calibrator
on and press [OK] on the analyzer.
13.1.5.2.2.
After calibration the analyzer will display the “Acoustic Sensor Calibration” screen.
13.1.5.3.
“Acoustic Sensor Calibration”
13.1.5.3.1.
The “Acoustic Sensor Calibration” screen provides details of the calibration. You will
have the option to change the calibration by pressing the [F1] “Change” key and
repeating the steps in paragraph 13.1.5.2.1. Otherwise press [OK] to complete the
calibration process and return to the “Select channel to calibrate” screen shown in