Setting-Up Tests
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
3-12
3
For the right eye, select SPHERE.
The trial lens cannot be calculated without you providing the patient’s
date of birth. If you have not entered the patient’s date of birth, enter
it at this screen by pressing DATE OF BIRTH.
Note: The Date of Birth button is disabled when recalling
patients from any source.
4
Enter the patient’s distance sphere correction. Always remember to
enter a plus (+) or minus (-) as the first character. Press ENTER.
If the patient has no sphere correction (plano), you must enter zero (0)
so that the proper trial lens calculation occurs.
5
Enter correction for cylinder and axis, if needed.
6
Repeat Steps 4-5 for the left eye.
Note: You may choose SPHERE, CYLINDER, and AXIS in any
sequence. To correct entries, reselect the command button and then
enter the correct data.
7
Select CALCULATE TRIAL LENS.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......