Kinetic Testing
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
13-16
4
Press AUTOMATIC.
You will then need to select the eye to test. The Patient Data screen
appears. Press the PROCEED button when all the desired patient data
has been entered.
5
Press START to begin the automatic test. The meridian to be tested
is indicated on the screen.
6
The stimulus will light but will not move for a short period of time.
You then will see a small diamond begin to move across the screen
and the STOP STIMULUS button will become active.
When the patient presses the response button, the isopter letter will
appear and the next meridian to be tested will display on the screen.
The HFA II-
i
will pause briefly, then start to move the stimulus. This is
the routine throughout the entire test.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......