Guided Progression Analysis (GPA)
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
8-20
GPA Case 3: Full Threshold as Baseline and Follow-up
Figure 8.9 Using Full Threshold Tests for a Baseline
This report shows an example of Full Threshold tests being used as the Baseline. When Full
Threshold tests are used for the Baseline, Full Threshold tests are eligible to be used as Follow-up
tests. In the VFI Plot open squares signify Full Threshold tests A. Closed squares signify SITA
Follow-up tests B. Remember that the Key Exam must always be a SITA Standard or SITA Fast test to
include Full Threshold tests in GPA.