Kinetic Testing
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
13-52
10
Here is an example of the second meridian added to the
program. The 80 degree meridian shows a gray box on the field. The
previous meridian (at 60 degrees) has changed to a circle. This is to
make it easier to identify the most recent meridian that you have
programmed.
11
When you have added all the meridians for the first isopter,
press the NEW ISOPTER button.
12
The Kinetic Parameter Setup screen will appear. Change the
stimulus parameters and the test speed to the desired settings for the
second isopter to be tested. Press SELECTION COMPLETE.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......