Patient B: This patient is about to be transferred to the local HFA database and has
the same Name and Date of Birth as Patient A, but does not have a Patient ID (i.e.
field is blank). When Patient B is copied, transferred, retrieved from an archive, or
restored from a backup to the local HFA database, Patient B will automatically be
assigned the same Patient ID and Issuer of ID as Patient A, and Patient B will be
merged with Patient A.
Transferring Tests from the HFA I to the HFA II-
i
Transferring results between the HFA I and HFA II-
i
is done via serial cable using the TRANSFER
TESTS function.
There are certain limitations to which tests will transfer from the HFA I to the HFA II-
i
. A name and
date of birth must be included on the original HFA I test in order to transfer it to the HFA II-
i
. The
transfer process will not warn you of improperly labeled HFA I tests or test results which are not
eligible to be transferred. The following is a list of restrictions specific to the test type being
transferred:
Threshold
• Fast Threshold or Master files will not transfer
• Compared, Averaged or Merged tests will not transfer
• Tests originally acquired on the HFA II-
i
(and transferred back from the HFA I) will not transfer
The Macula test: Only the first two threshold values at each point will be displayed and printed.
Note: The Peripheral 30/60-2 test on the HFA I is the Peripheral 60-4 on the HFA II-
i
. The test results
will transfer, but the 4 points not found in the 60-4 pattern will not be included.
Screening
All screening tests will transfer from the HFA I to HFA II-
i
except:
• The C-166 Screening test
• A screening test where the user selected the Central or Peripheral Reference Level
• Auto Diagnostic tests
Note: The Peripheral 68 test on the HFA I is the P-60 on the HFA II-
i
. These test results will transfer,
but the 8 points not found in the P-60 pattern will not be included.
Custom
Custom Arc and Profile tests will not transfer.
Kinetic
Kinetic test results cannot be transferred from the HFA I.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......