Kinetic Testing
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
13-36
6
When you have located that position, press SELECT END POINT.
The stimulus will begin moving when you activate SELECT END POINT.
During the test a small diamond moves across the video display. It dis-
plays the progress of the stimulus in the test bowl.
7
To perform another Custom Scan, repeat the previous steps.
Continue until you have determined all desired points for each
particular isopter.
After you have mapped all desired points, press END CUSTOM SCAN
to return to the Kinetic Test screen.
8
If all of the desired stimuli from the different isopters have been
tested, press END TEST from the Kinetic Test screen. In the illustration,
you can see an example of a Custom Scan that has added definition to
the inferior border of a hemianopsia.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......