Care and Cleaning
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
15-8
Replacing Power Table Fuses
1
Turn off the power. Unplug the power cord that is attached to the table at the base of the
lifting column. Move the table away from the wall to access the back of the table.
2
Use needle-nose pliers or fingernails gently to slide the fuse drawer from its housing.
3
Remove defective fuses. Insert new fuses in housing.
4
Replace the fuse drawer. Plug in the power cord.
Note: For continued protection against fire hazard, replace only with same type and
rating of fuse.