Setting-Up Tests
3-23
Inserting Trial Lenses Into the Holder
1
Move the trial lens holder into an upright position from its storage
position in the bottom of the bowl.
2
Place the cylinder lens in the slot farthest away from the patient
and align the axis.
3
Place the sphere lens in the slot closest to the patient (in front of
the cylinder lens).
Note: Use only the narrow rimmed type of trial lenses. The wide-rimmed variety will interfere with
the patient’s peripheral vision and adversely affect test results. It is helpful to move the lens handle
towards the patient’s temporal side so it does not interfere with the patient’s eye brow or nose.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......