Custom Testing
Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
12-6
6
The Custom screen will appear.
Your chosen Point Pattern, Field Size, and Point Spacing are indicated
in the lower, right-hand corner of the screen.
Note: Custom tests are automatically generated for the right
eye only. When the left eye is tested, the HFA II-
i
will flip the test
pattern to accommodate for the mirrored physiological differences. Do
not attempt to design a left eye Custom test.
Adding Single Points
1
If you chose the Single Point (X, Y) option, you may now input the individual points to be
tested. Press the ADD ENTRY button.
2
For a Single Point entry, you must enter the (X, Y) coordinate values
(as whole numbers) of each point you wish to test.
Input the “X” value of the point. Press ENTER.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......