Humphrey Field Analyzer II-
i
series User Manual
2660021145640 A
Short-Wavelength Automated Perimetry (SWAP)
9-7
4. As with white-on-white perimetry, we recommend that you confirm any abnormal findings
before making a diagnosis that is based on the visual field results. Patients often exhibit a
“learning effect” and may be better prepared for the SITA-SWAP testing routine on a second
visit. This leads to an improved recognition of the blue stimulus, as well as better pacing of
patient responses. These factors combine to provide more accurate test results.
5. It is possible to use the Full Threshold and FastPac testing strategies to do SWAP. However, we
do not recommend that you use these strategies, as the test times can be two or three times
longer than those of a SITA-SWAP test. Further information on Full Threshold and FastPac test
strategies is presented in
.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......