Short-Wavelength Automated Perimetry (SWAP)
9-11
Case 2: Comparing a SITA-SWAP Test with Standard White-on-White Testing
Figure 9.7a Glaucomatous Visual Field Defect Discovered with SITA-SWAP
In this case, an early paracentral visual field defect is seen with SITA-SWAP in the superior hemifield
of the patient's right eye. The GHT indicates “Outside Normal Limits”. Both the MD and PSD global
indices are flagged, with the PSD marked at the P < 5% limit.
This printout is a good example of why you should avoid referring to the grey tone diagram with
SWAP testing. A larger area of the superior visual field appears to be affected. Yet, the Total and
Pattern Deviation probability plots show a much smaller but important area to be significantly
depressed relative to normal limits.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......