Acquisition and Display
7-27
IM 701310-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
App
Index
Explanation
The DL9000 returns GO/NO-GO results based on whether or not the source waveform
falls within the zone that you set using a reference waveform.
Reference Waveform
Select the waveform that you will use as a basis for the GO/NO-GO determination zone.
You can select the reference waveform from CH1 to CH4 or from M1 to M4. You cannot
select a waveform that is not displayed.
Creating a GO/NO-GO Determination Zone
You can create a maximum of four GO/NO-GO determination zones. The boundaries of
the zones can be set within the following ranges:
• Upper and lower boundaries: �8 divisions from the reference waveform
• Left and right boundaries: �5 divisions from the center of the screen
You can choose to apply the GO/NO-GO determination zones assigned to GO/NO-
GO condition numbers 1 to 4 to input signal waveforms (CH1 to CH4) and computed
waveforms (M1 to M4). Any zone whose condition setting is either IN or OUT appears on
the display.
Source Waveform
You can choose a waveform from CH1 to CH4 or from M1 to M4.
GO/NO-GO Condition Number
You can choose to configure a GO/NO-GO condition from 1 to 4.
GO/NO-GO Determination Area
You can set the GO/NO-GO determination area to less than or equal to �5 divisions from
the time axis.
GO/NO-GO Determination Condition
• IN: The DL9000 returns a no-go result when the source waveform is within the GO/
NO-GO determination zone.
• OUT: The DL9000 returns a no-go result when the source waveform is outside of the
GO/NO-GO determination zone.
• X: GO/NO-GO determination does not take place.