Acquisition and Display
7-19
IM 701310-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
App
Index
• E-Mail Transmission
The DL9000 sends an e-mail to the specified address (if it is equipped with the
Ethernet interface option).
To set the e-mail address, see section 15.4, “Configuring E-Mail Transmission.”
Operations Performed for the Print and Save to File Actions
Operations are performed in accordance with the settings made in the PRINT and FILE
menus. To access these menus and their settings, see chapter 12, “Printing Screen
Images,” section 13.5, “Saving/Loading the Measurement Data,” and section 13.8, “Saving
Screen Image Data.” When the File Menu’s automatic naming function is OFF, file names
will be assigned through numbering. If automatic naming is not OFF, filenames will be
assigned according to the specified method.
Mail Transmission Operations
Mail-Mode ON/OFF
When Mode is ON, e-mails will be sent to the address specified in: SYSTEM menu >
Network > E-mail Setup > Address (To Address/From Address).
Transmit Interval
The DL9000 will send an e-mail at the first trigger activation after the interval specified
here elapses. An e-mail will also be sent when GO-NOGO determination or the action-
on-trigger function stops (stopping includes manual stops).
The mail transmission interval can be set to a value from OFF to 1440 min, in 1-min
steps. If you select OFF, an e-mail will be sent whenever a trigger is activated.
• An example of e-mail transmission with a set interval.
Interval
Action-on-trigger/determination start
Interval
Interval
Action-on-trigger/determination stop
(including manual)
Trigger
occurrence
Trigger occurrence
and
mail transmission
Trigger occurrence
and
mail transmission
Mail transmission
Trigger
occurrence
Trigger
occurrence
E-Mail Content
• When Nogo Is Set as an Action and Interval Is Set to OFF
<Subject>:
The subject of the e-mail. The content of the subject is as
follows:
GoNogo Triggered Report (No.).
The number in parentheses is the number of returned no-go
results.
[Comment]:
Comments
[Setup Information]:
Parameter and maximum and minimum value information
about GO/NO-GO conditions 1 to 4.
Logic (AND/OR)
Stop Nogo/ACQ Count (The number of no-go results/the
acquisition count)
[Trigger Date and Time]: The trigger date and time
[Nogo/Exec Count]:
The number of no-go results/the number of determinations
[Nogo Factor]:
Information, including measured values, about the conditions
that returned no-go results*
* Measured values are only sent for GO/NO-GO determination based on waveform
parameters.
7.9 Activating the Action-On-Trigger Function Using GO/NO-GO Results