19-11
IM 701310-01E
Specifications
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
App
Index
19.11 General Specifications
Item
Specifications
Standard operating conditions Ambient temperature:
23 ±5˚C
Ambient humidity:
55 �10%RH
Power supply voltage and frequency tolerance: Within 1% of rated value
Warm-up time
At least 30 minutes
Storage conditions
Temperature:
–20 to 60˚C
Humidity:
20 to 80%RH (no condensation)
Operating conditions
Temperature:
5 to 40˚C
Humidity:
20 to 80%RH (not using printer) (no condensation)
35 to 80%RH (using printer) (no condensation)
Recommended calibration
period
1 year
Storage altitude
3000 m or less
Operating altitude
2000 m or less
Rated supply voltage
100 to 120 VAC or 220 to 240 VAC (automatic switching)
Power supply voltage
90 to 132 VAC or 198 to 264 VAC
fluctuation tolerance
Rated supply frequency
50/60 Hz
Power supply frequency
48 to 63 Hz
fluctuation tolerance
Power supply fuse
Internal (not replaceable)
Maximum power
consumption
300 VA
Withstand voltage (between 1.5 kVAC, one minute
power supply and case)
Insulation resistance
(between power supply and
case)
500 VDC, at least 10 M
W
External dimensions
350(W)
×
200(H)
×
178(D) mm (with printer cover packed away, excluding handle and other
projections)
Weight
Approx. 6.5 kg (including printer)
Cooling method
Forced air cooling, side ventilation
Installation orientation
Horizontal (stand or handle may be used). Vertical and stacking prohibited.
Battery backup
Setup data and clock are backed up with the internal lithium battery