2-28
IM AQ7280-01EN
Cable Installation Completion Notification
1.
On the OTDR menu, press the
Advanced Measurement
soft key and then the
Work
Completion Notice
soft key. The following screen appears.
Turns the completion notification
on or off
Turns the message display on or off
Turns the alarm on or off
Set the confirmation (No Check,
Check only once, Check everytime).
2.
Set Notification to
ON
.
3.
Set Message to
ON
.
4.
Press
REAL TIME
to start a measurement. The screen shows the measurement waveform, and
the RealTime menu appears. The words “Check Point” appears at the cursor on the screen.
5.
Press the
Next
soft key and then the
Completion Check
soft key to select ON.
When the measurement to the Check Point distance is complete, a completion message is displayed.
Set the work completion
confirmation (ON, OFF).
When the work completion notice is set to ON,
“CheckPoint” appears at the cursor.
End of fiber
Move the cursor to a
position that is slightly
after the installation
point.
You can perform measurements efficiently by moving the cursor to a position
slightly after the installation point (depending on how fast the optical fiber cable will
be installed). This is because the AQ7280 notifies you of the work completion, which
is your sign to measure the connection loss at that point (splice, connection, etc.).
2.5 Performing Real-time Measurement