4 - 41
Detector Data Setup
Introduction
This is used to set and select the constant, model, and diameter of the detector
to be used in combination with the converter.
Default setting
When there is no detector setting, EX300.0, MGG, DIA 050.0 will be se-
lected.
Note:
• When you purchase the converter and detector in combination,
your converter will contain the detector data that was set during
actual flow calibration. Take care not to change the data, or the
flowmeter output will be incorrect. Refer to Table 4-1.
1
Open the detector data setup
screen by following the steps to
enter the Engineering Mode.
2
Touch the key to set the
detector constant. Use the
and keys to input the
numerical value printed on the
EX column of the detector
nameplate.
3
Touch the key to select the
type of detector. Use the
and keys to select the
model number printed on the
nameplate of the detector to be
used.
4
Touch the key to select the
diameter.
Use the and keys to select
the diameter of the detector to
be used.
5
Use the key to move the
cursor to the "#".
Screen
Procedure
Step
Continued on next page
# EX 300.0
MGG DIA 050.0
# EX 320.0
MGG DIA 050.0
# EX 320.0
KID DIA 050.0
# EX 320.0
KID DIA 100.0
# EX 320.0
KID DIA 100.0