3 - 2
3.1 Start-up
Starting up
Procedures
Start up the electromagnetic flowmeter according to the following steps.
Step
Procedure
1
Make sure the electromagnetic flowmeter detector is prop-
erly installed on the pipe.
2
Make sure the wiring between the electromagnetic flowmeter
detector and converter has been completed properly. For
communication with the S-SFC, make
sure the S-SFC wiring has also been properly completed.
3
Charge the electromagnetic flowmeter with the fluid to be
measured, and make the fluid stand still.
Make sure there is no fluid leaking from the flange to which
the electromagnetic flowmeter is attached.
4
5
Turn the power to the electromagnetic flowmeter ON.
Branch:
• For flowmeters without the data setting device, this is the
completion of start-up. Next, start up the S-SFC.
• For flowmeters with the data setting device, proceed to
Step 6.
6
Make a display similar to the one shown below appears on the
LCD.
Start-up of the electromagnetic flowmeter has now been com-
pleted.
0.00 m 3/h
0000123456 TOTAL