5 - 13
Error Messages and Treatment (Display)
Hardware check
Hardware check is executed in the Measuring Mode.
Perform the proper measures immediately.
While an error code (Err-01 to Err-05) is being displayed, the output for the
abnormal situation will be as preset in the Engineering Mode.
Error codes for
serious trouble
EX CHECK
ERROR
E r r - 01
1. Check connection.
2. Measure coil resistance.
3. Restore power.
Coil disconnected
EX open
Err
–
01
Error
Error contents
Treatment
LCD display
code
1. Restore power.
2. Replace ROM.
3. Replace main circuit
board.
ROM
check
error
Err
–
02
1. Restore power.
2. Replace main circuit
board.
RAM
READ AFTER
WRITE error
Err
–
03
1. Restore power.
2. Replace main circuit
board.
NVM
READ AFTER
WRITE error
Err
–
04
1. Restore power.
2. Replace main circuit
board.
ADC error
A/D change error
Err
–
05
(Continued on next page)
EX CHECK
ERROR
ROM CHECK
ERROR
RAM CHECK
ERROR
NVM CHECK
ERROR
ADC CHECK
ERROR